Reassembling The Nvaei Ion Source Cartridge - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

Figure 68. Components of the NeverVent Advanced EI Ion Source Cartridge
NVAEI Cap
NVAEI Filament
Spring

Reassembling the NVAEI Ion Source Cartridge

Thermo Scientific
NVAEI Filament
Insulator
NVAEI
Filament
AEI Filament
Shield
2. Set the components on a clean work surface.
3. Separate the repeller nut, ion volume-repeller insulator, and repeller.
Clean the ion source cartridge as instructed in
1. To reassemble the NVAEI ion source cartridge, begin by inserting lens 3 into the ion
cartridge sleeve. Align the grooves on lens 3 with the ones on the sleeve.
Lens 3/RF
Lens
AEI Ion Volume
Insulator
Lens 2
NVAEI Ion Volume
Cleaning Durable
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Ion Cartridge Sleeve
Components.
TSQ 9610 Hardware Manual