Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
70
TSQ 9610 Hardware Manual
Figure 69. Inserting Lens 3 into the Ion Cartridge Sleeve
2. Insert lens 2 with the flat side down into the ion cartridge sleeve.
Figure 70. Inserting Lens 2 into the Ion Cartridge Sleeve
3. Attach the ion volume insulator to the AEI ion volume. A notch on the side of the
insulator lines up with a key on the ion volume.
Figure 71. Attaching the Ion Volume Insulator to the AEI Ion Volume
4. Insert the AEI ion volume and insulator into the ion cartridge sleeve.
Thermo Scientific