Thermo Scientific
18. Let the ion source cartridge cool.
19. Hold the handle of the source exchange tool with one hand and use the other hand to
pull the barrel toward you and into the handle.
Figure 46. Exposing the Ion Source Cartridge
20. Invert the source exchange tool so that the barrel is pointed toward the floor.
Figure 47. Inverting the Tool
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Ion Source
Cartridge
Ion Source
Cartridge
TSQ 9610 Hardware Manual
55