Figure 117. Disconnecting the Ion Guide Connector Wires and the Quad 1 Connector Wires
Front 4-pin
Connector
Ion Guide Wires
6. Disconnect the four wires that go to the collision cell and quadrupole 3. These wires are
attached to the rear 4-pin feedthrough in the vacuum manifold.
Figure 118. Disconnecting the Quad 3 Connector Wires and the Collision Cell Connector Wires
Rear 4-pin
Connector
Collision Cell
Connector Wires
7. Remove the collision gas tube from the collision cell.
Note Pull straight back from the collision cell to prevent damage to the collision gas tube.
8. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Quad 1
Wires
Quad 3
Wires
Collision Gas
Tube
TSQ 9610 Hardware Manual