Advanced Troubleshooting
Replacing Components of the Analyzer
150
TSQ 9610 Hardware Manual
9. Set the collision cell assembly on a clean work surface. Disconnect the Q1 entrance lens
wire from the source interface board.
Figure 129. Removing the Collision Cell and Q1 Quad Assembly from the Analyzer Tray
Collision
Cell
Assembly
Screw to
Analyzer Tray
Figure 130. Removing the Analyzer Tray Screws
Ion Guide Screws
10. Remove the Q1 entrance lens wire from the source interface board.
11. Remove the ion guide assembly screws.
12. Remove the ion guide assembly.
13. Use a T10 Torxhead screwdriver to remove the lens heater block grounding strap from the
analyzer tray.
Q1 Entrance Lens
Screw to
Endcap Screws to
Axial Field
Analyzer Tray
Wire
Q3 Entrance Endcap Screw
Q1 Exit Endcap
Screw
Q1 Quad
Assembly
Q1 Entrance
Lens Endcap
Q3 Exit Endcap
Screw