Thermo Scientific
3. Disconnect the vacuum interlock cable from the connector.
Figure 301. Disconnecting the Vacuum Interlock Cable
4. Pinch together the two springs on the left side the vacuum interlock shield and remove it.
Figure 302. Removing the Vacuum Interlock Shield
Springs
5. Lift the vacuum interlock shield off the instrument. The notch at the bottom of the track
will go around the handle. Be sure to straighten the ball valve handle prior to removing
the VPI shield.
3
Advanced Troubleshooting
Replacing Components of the Manifold
Vacuum
Interlock
Cable
Connector
Ball Valve
Handle
TSQ 9610 Hardware Manual
273