Turbo Cooling Fan and Turbo Cooling Fan Filter

Thermo Scientific
To replace the turbo cooling fan and turbo cooling fan filter
Note See the ISQ and TSQ GC-MS Spare Parts Guide for ordering information.
1. Shut down your system by following the steps in
on
page
139. You do not need to remove the ion source cartridge.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Move the GC away from the TSQ 9610 instrument.
4. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
Figure 454. Removing the Right Side Panel Screws
5. Lift the right side panel off the instrument.
6. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
"Shutting Down the TSQ 9610 System"
Right Side
Panel Screws
TSQ 9610 Hardware Manual
3
Advanced Troubleshooting
Replacing Fans and Filters
415