Thermo Scientific TSQ 9610 Hardware Manual page 172

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Analyzer
142
TSQ 9610 Hardware Manual
2. Disconnect the electrometer board cable
Figure 115. Disconnecting the Electrometer Board Cable
Electrometer
Board Cable
3. Remove the small and large glass manifold cover.
4. Set the glass covers on a surface where they will not be scratched or contaminated. Protect
the glass and any other manifold components from dust accumulation. Wear cleanroom
gloves when touching the glass. See the ISQ and TSQ GC-MS Spare Parts Guide for
information about ordering cleanroom gloves.
Figure 116. Removing the Small and Large Glass Manifold Covers
5. Disconnect the four wires that go to the ion guide and quadrupole 1. These wires are
attached to the front 4-pin feedthrough in the vacuum manifold.
Large Glass
Manifold Cover
Small Glass
Manifold Cover
Thermo Scientific