• Busy—If the Busy light is a solid blue, the instrument is actively scanning. When the
light is not lit, the instrument is not scanning. The Busy light is also lit when
establishing communication with the computer or when the filament or electron
multiplier is on.
6. If you are running Xcalibur software, open the TSQ 9610 Dashboard by clicking on the
desktop icon.
7. Check the status of the vacuum on the TSQ 9610 Dashboard. Within 10 minutes of
powering on the instrument, the vacuum should read OK.
8. Check the heater status on the TSQ 9610 Dashboard. If the ion source is not set to the
desired temperature, click the Instrument Control button and change the
temperature.The temperature set point will change when the send button is pressed.
Figure 4.
Checking the Status of TSQ 9610 System in Xcalibur Software
9. If you are running Chromeleon software, check the instrument status in the
Detector-Service utility. See
Figure
5.
1
Performing Routine Maintenance
Powering On the TSQ 9610 System