3
Advanced Troubleshooting
Replacing Components of the Analyzer
Replacing the Analyzer Tray in the TSQ 9610 Instrument
154
TSQ 9610 Hardware Manual
IMPORTANT Be sure that the ceramics are aligned correctly before fastening the endcap
to the analyzer tray.
13. Reconnect the three lens wires attached to the lenses between the quadrupoles.
14. Reconnect the Q1 entrance and Q1 exit lens assemblies.
15. Reconnect the Q1 entrance lens wire to the source interface board.
To replace the analyzer tray in the TSQ 9610 instrument
1. Put tray in manifold with transfer line out or pulled away.
a. If the gas tubes are attached to the transfer line, it will not pull as far away from the
path of the tray as would be best. Be extra careful not to scratch the end of
transferline with the tray, because that could cause a leak with the source plug later.
b. Do not screw in the analyzer tray until the transfer line is fully screwed in.
Note The analyzer tray will mount on one alignment pin in the back and the transfer line
on the right. The mixing chamber will rest on the transfer line when installed.
Figure 135. Sliding the Analyzer Tray Back into the Instrument
CAUTION INSTRUMENT DAMAGE Do not scratch any of the sealing surfaces. Be
especially careful when the source interface board reaches the o-ring surface.
Sealing Surfaces
Analyzer Tray
Tab
Thermo Scientific