3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument

Electrometer Board Cable

366
TSQ 9610 Hardware Manual
To replace the electrometer board (15-pin male/female RS-232) cable
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139. You do not have to remove the ion source cartridge or the column or vent
the instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Disconnect the electrometer board cable
Figure 406. Disconnecting the Electrometer Board Cable
Electrometer
Board Cable
4. Remove all power connections from the TSQ 9610 instrument.
"Shutting Down the TSQ 9610 System"
Thermo Scientific