Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
362
TSQ 9610 Hardware Manual
43. Reconnect the large feedthrough cables on the electron multiplier plate.
44. Reattach the RF boards by aligning the connectors on the bottom of the board with the
connectors on the distribution board beneath it. The boards will drop into their slots
when they are in the right position.
45. Gently push the RF board into the instrument, carefully aligning the board over the
feedthrough pins, until it stops.
46. Reattach the screw holding the RF board to the bracket.
47. Reinstall the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will click into place when its in the
right position.
Figure 401. Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board
48. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
tighten the screw holding the lens driver board in place on the chassis.Reattach the
controller interface board by aligning the connectors on the bottom of the board with the