Thermo Scientific TSQ 9610 Hardware Manual page 413

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
7. Remove the controller interface board from the instrument.
8. Look through the top of the instrument and locate both RF boards. The front RF board
is the second board standing on end toward the center of the instrument. The rear RF
board is the third board standing on end toward the center of the TSQ 9610 instrument.
Use a T20 Torxhead screwdriver to remove the screw holding the board and bracket to
the chassis.
Figure 421. Locating the RF Boards and Brackets
.
9. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument. Lift the RF boards out of the instrument and set them aside.
Replacing the Calibration Gas Components
Rear RF
Front RF
Board
Board
Brackets
TSQ 9610 Hardware Manual
3
Advanced Troubleshooting
383

Advertisement

Table of Contents
loading

Table of Contents