9. Reattach the thumbscrew-repeller spacers.
Note When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crack the spacers.
10. Reattach the source-repeller thumbscrews to the spacers.
11. Reattach the manifold door and the four screws.
12. Reattach the top cover and tighten the top cover panel screw.
13. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing Components of the Ion Source
"Restarting the System"