7. Use a T20 Torxhead screwdriver to remove the five screws on the right side panel.
Figure 351. Removing the Right Side Panel Screws
8. Slide the right side panel to the right of the instrument and lift it off.
9. Use a 5.5 mm open head wrench to hold the M-3 lock washers and remove the two M3
flathead screws suing a T10 Torxhead screwdriver.
Figure 352. Removing the Latch from the Chassis
10. Attach a new latch to the chassis.
11. Reattach the right side and top cover panels and tighten the screws.
12. Reattach the GC to the transfer line.
13. Restart your system by following the instructions in