Checking The Oil In The Foreline Pump; Adding Oil To The Foreline Pump - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

1
Performing Routine Maintenance
Maintaining the Foreline Pump

Checking the Oil in the Foreline Pump

Adding Oil to the Foreline Pump

36
TSQ 9610 Hardware Manual
To check the oil level of the foreline pump:
1. Ensure the pump is on a level surface.
2. Look in the oil-level sight-glass to see if the oil level is between the MIN and MAX marks.
3. If the oil level is below the MIN mark, add oil, as described in
Pump.
4. Look for oil that is clear or light orange in color.
a. If the oil is cloudy or discolored, purge the oil to decontaminate dissolved solvents, as
described in
Purging Gas from the Oil in the Foreline
b. If the pump oil is still discolored, change it, as described in
Foreline
Pump.
To add oil to the foreline pump:.
Note If you are running samples, stop the acquisition before powering off the system.
CAUTION BURN HAZARD: The oven may be hot. Allow it to cool to room temperature
before opening it. The injector will still be hot, so do not touch it.
ATTENTION RISQUE DE BRÛLURE : le four peut être chaud. Laissez-le refroidir à la
température ambiante avant de l'ouvrir. L'injecteur sera toujours chaud, donc ne le touchez
pas.
WARNING FIRE HAZARD: If you are using hydrogen as a carrier gas, you must cool down
and shut off the GC to prevent the buildup of hydrogen in the vacuum manifold. See the
"Hydrogen Safety Precautions"
AVERTISSEMENT RISQUE D'INCENDIE : si vous utilisez de l'hydrogène comme gaz
vecteur, vous devez refroidir et mettre hors tension le CG pour éviter l'accumulation
d'hydrogène dans le collecteur à vide. Reportez-vous aux « Précautions de sécurité relatives à
l'hydrogène » à la page xix de la préface pour plus d'informations.
1. Click the Shut Down button on the TSQ 9610 Dashboard.
2. Click the Yes button to continue the shutdown process. The heaters and turbomolecular
pump power off. Once the turbomolecular pump reaches 50% rpm, the foreline pump
on
page xxi
of the Preface for more information.
Adding Oil to the Foreline
Pump.
Changing the Oil in a