9. Slide the repeller spring onto the repeller.
Figure 66. Inserting the Repeller Spring into the Sleeve
10. Place the locking ring on top of the repeller spring so that the repeller protrudes through
the center hole on the locking ring. The hooks on the sleeve fit between the larger gaps on
the locking ring.
Note Do not twist and lock the locking ring on the sleeve at this time.
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Repeller
Spring
TSQ 9610 Hardware Manual