Thermo Scientific TSQ 9610 Hardware Manual page 461

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
9. Insert and twist the convectron gauge into the threaded hole on the foreline adapter.
Note Before you insert the convectron gauge in the hole, wrap Teflon tape around the
part to be inserted. The tape prevents vacuum leaks.
Figure 471. Installing the Foreline Adapter to the Convectron Gauge
Tape
Foreline Adapter
10. Attach the foreline hose on the adapter.
11. Use the rough pump clamp to attach the adapter to the turbo pump.
12. Attach the cable to the distribution board and convectron gauge.
13. Reattach the right side panel and the screws holding it in place.
14. Replace the top cover panel and tighten the screw.
15. Reattach the GC.
16. Restart your system by following the instructions in
3
Advanced Troubleshooting
Convectron Gauge and Foreline Adapter with Hose
Convectron Gauge
"Restarting the System"
TSQ 9610 Hardware Manual
on
page
158.
431

Advertisement

Table of Contents
loading

Table of Contents