Thermo Scientific
Figure 80. Attaching the NVAEI Ion Source to the NVAEI Source Exchange Tool
6. Gently pull the source holder away from the ion source cartridge.
Figure 81. Removing the Source Holder
Tip When the source exchange tool is inverted, the source should stay attached. If this
does not happen, the source will not insert correctly into the instrument.
7. Turn the source exchange tool around so that the ion source cartridge is furthest away
from you.
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
TSQ 9610 Hardware Manual
75