Thermo Scientific TSQ 9610 Hardware Manual page 55

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
the air has been evacuated from the instrument and it is safe to open the vacuum interlock
valve.
Note If the vacuum light flashes amber, there is a leak in the system.
10. Pull the vacuum interlock handle up when the Ready to Open light is a solid green.
11. Twist the handle of the tool slightly to the left until it is lodged into the left-most track.
See
Figure
23.
Figure 23. Inserting the V-lock Source Plug
12. Push the handle toward the instrument until the end of the handle aligns with the
engraved line at the end of the barrel. When you reach this line, the tool is all the way in
and the V-lock source plug is securely placed in the instrument. See
IMPORTANT The V-lock source plug should remain attached to the source exchange tool.
Do not rotate the source exchange tool handle to attempt to disengage the source plug.
Figure 24. V-lock Source Plug Inserted Correctly
1
Performing Routine Maintenance
Replacing a Column
Figure
24.
TSQ 9610 Hardware Manual
25

Advertisement

Table of Contents
loading

Table of Contents