Thermo Scientific TSQ 9610 Hardware Manual page 482

Mass spectrometer
Table of Contents

Advertisement

4
Upgrade Equipment
Ion Gauge and Tube Shield
452
TSQ 9610 Hardware Manual
Figure 498. Assembling the Ion Gauge
Board
Shield
7. Use a T10 Torxhead screwdriver to attach the screws holding the metal ion gauge shield
to the board.
8. Attach the glass ion gauge tube to the shield so the open end is facing away from the
board. There is only one way to fit the tube's electrical connectors onto the board.
Tip Insert the tube until there is good electrical contact and ample room on the open end
for the o-ring to fit securely. This will prevent the o-ring from sliding off when the
assembly is being put on the manifold.
9. Attach the 14-pin cable to the distribution board and ion gauge board.
Figure 499. Connecting the Ion Gauge Cable to the Distribution Board
10. Remove the plug on the bottom of the manifold.
Ion Gauge
Cable
Connector
Screws (3)
Tube
O-Ring
Thermo Scientific

Advertisement

Table of Contents
loading

Table of Contents