Thermo Scientific TSQ 9610 Hardware Manual page 430

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Collision Gas Components
400
TSQ 9610 Hardware Manual
Figure 439. Locating the Collision Gas Module
20. Remove the large glass manifold cover.
Note You may skip this step if you are not removing the collision gas tubing from the
collision cell.
21. If you are replacing the collision gas module and do not need to replace the tubing to the
collision cell, you may leave the tubing attached to the manifold and collision cell and
remove the tubing from the collision gas module only. If you are replacing the tubing to
the collision cell, disconnect the collision gas tubing from the collision cell and from the
compression fitting on the manifold.
Figure 440. Disconnecting the Tubing from the Collision Cell
Tubing to
Foreline Adapter
Collision Cell
Collision Gas
Module
Collision Gas
Tubing
Thermo Scientific