Thermo Scientific TSQ 9610 Hardware Manual page 292

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
262
TSQ 9610 Hardware Manual
32. Reattach the three screws holding the metal valve body to the front manifold plate.
Figure 288. Reattaching the Metal Valve Body to the Front Manifold Plate
Screws
33. Reattach the two screws connecting the black pillow block to the manifold door.
Figure 289. Reattaching the Clamp
Black Pillow Block
Screws
CAUTION INSTRUMENT DAMAGE: If the vacuum interlock handle is not fully down
and the ball valve is not fully closed, then the ball valve may not be able to completely
open during use and it could damage the ball valve and probes. Confirm the ball is fully
opened and closed as the handle is moved up and down.
Front
Manifold
Plate