22. Put the electron multiplier plate back into place and reattach the four screws. Tighten
them in this order: screw 1, screw 2, screw 3, screw 4. See
Figure 101. Reattaching the Electron Multiplier Plate
Screw 1
Screw 4
23. Reconnect the electrometer board cable and large feedthrough cables.
CAUTION INSTRUMENT DAMAGE: Do not mix up the cables or you will damage the
power supply.
ATTENTION DOMMAGES À L'INSTRUMENT : ne confondez pas les câbles ou vous
endommagerez l'alimentation.
1
Performing Routine Maintenance
Replacing the Electron Multiplier
Figure 101
for clarification.
Screw 3
Screw 2
TSQ 9610 Hardware Manual
97