Thermo Scientific TSQ 9610 Hardware Manual page 321

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
6. Use a T10 Torxhead screwdriver to remove the anode feedthrough, two screws, standoff.
and o-ring. Set them aside on a clean work surface.
Figure 328. Removing the Anode Feedthrough
Anode
Feedthrough
and O-ring
7. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
Figure 329. Disconnecting the Electron Multiplier Plate
Electron Multiplier
Plate Screws
3
Advanced Troubleshooting
Replacing Components of the Electron Multiplier
Standoff
TSQ 9610 Hardware Manual
Screws
291