Thermo Scientific TSQ 9610 Hardware Manual page 365

Mass spectrometer
Table of Contents

Advertisement

Thermo Scientific
9. Insert a new lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
Figure 370. Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board
10. Push the board down into the connector.
11. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
tighten the screw holding the lens driver board in place on the chassis.
12. Reattach the top cover and left side panels and tighten the screws.
13. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
"Restarting the System"
TSQ 9610 Hardware Manual
on
page
158.