Thumbscrew-Repeller Spacers - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

Thumbscrew-Repeller Spacers

Thermo Scientific
To replace the thumbscrew-repeller spacers
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
3. Open the manifold door.
Figure 187. Opening the Manifold Door
3
Advanced Troubleshooting
Replacing Components of the Ion Source
"Shutting Down the TSQ 9610 System"
Screws
TSQ 9610 Hardware Manual
193