Thermo Scientific TSQ 9610 Hardware Manual page 196

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Analyzer
166
TSQ 9610 Hardware Manual
Figure 149. Removing the Analyzer Tray Screws
Ion Guide Screws
Figure 150. Removing the Collision Cell and Quadrupole 3 from the Analyzer Tray
Q3 Quad
Assembly
Collision Cell
Assembly
8. Disconnect the Q1 entrance lens wire from the source interface board.
9. Remove the two screws on each side of the Q1 entrance endcap.
Figure 151. Removing Screws from the Quadrupole Rod Endcaps
Q1 Entrance
Lens and Endcap
Screws
Q3 Entrance Endcap Screw
Q1 Exit Endcap
Screw
Q3 Exit Endcap
Screw
Q3 Exit Lens
Assembly
Screws
Collision Cell