ATTENTION DOMMAGES À L'INSTRUMENT Limitez au maximum la manipulation des
cartes. Ne touchez aucun composant de carte. Gardez la carte uniquement près du support
en tôle.
.
Figure 372. Locating the Lens Driver Board
6. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the board to the chassis. Loosening the screw will separate the
board from the instrument. The screw is captured, so do not attempt to completely
remove the screw from the board and bracket.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately two inches from the top of the board directly under the white label.
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
Lens Driver