3
Advanced Troubleshooting
Replacing Components of the Ion Source
192
TSQ 9610 Hardware Manual
5. Extract the thumbscrew-repeller spacers and transfer to new thumbscrews.
Figure 186. Replacing the Source-Repeller Thumbscrews
6. Attach new source-repeller thumbscrews to the ion source.
7. Reattach the manifold door and the four screws.
8. Restart your system by following the instructions in
Source-Repeller
Thumbscrews
"Restarting the System"
on
Thermo Scientific
page
158.