Figure 145. Reattaching Quadrupole 1, the Ion Guide, and the Collision Cell Assembly to the
Analyzer Tray
Collision Cell
.
Figure 146. Reattaching Quadrupole 1, the Ion Guide, and the Collision Cell Assembly to the
Analyzer Tray
Q1 Entrance
Lens Endcap
Tip You will get better signal if you gently push the quadrupole rod endcaps toward one
another while tightening the screws.
22. Reattach the Q3 exit lens assembly and Q3 to the analyzer tray. Ensure that quadrupole 3
and the collision cell assembly are correctly positioned before tightening screws.
23. Reattach the two screws on each side of the quadrupole rod endcaps.
Replacing Components of the Analyzer
Screws
Screws
TSQ 9610 Hardware Manual
3
Advanced Troubleshooting
Q1 Quad
Assembly
Ion Guide
Quadrupole 1
Collision Cell
Assembly
163