8. Slide the source interface board and source block onto the tray without damaging or
disconnecting the wires.
Figure 206. Sliding the Ion Source Block onto the Analyzer Tray
Wires
9. Reattach the two screws to the back of the source block.
Figure 207. Reconnecting the Ion Source to the Analyzer Tray
Screws
IMPORTANT When assembling the source and optics block, take extra care to ensure the
blocks are centered properly and that the ion source cartridge provided with the
instrument can easily slide in and out of the assembly.
3
Advanced Troubleshooting
Replacing Components of the Ion Source
Source
Interface
Board
Source Block