6. Remove the four screws holding the electrometer board in place
Figure 409. Replacing the Electrometer Board
Screws
7. Remove and replace the electrometer board if necessary.
8. Reattach the screws that hold the electrometer board in place.
9. Reattach the electrometer shield and screws.
Tip To prevent the screws from falling down into the instrument, place them on the shield
before you reattach it to the chassis.
10. Reconnect the electrometer board cable.
11. Reattach the top cover panel and tighten the screw holding it in place.
12. Restart your system by following the instructions in
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
"Restarting the System"
TSQ 9610 Hardware Manual
on
page
158.