Advanced Troubleshooting
Replacing Components of the Electron Multiplier

Large Feedthroughs

294
TSQ 9610 Hardware Manual
To replace the large feedthroughs, o-rings, and screws
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
on
page
139.
Note You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Disconnect the electrometer board cable and large feedthrough cables labeled Dynode
and Electron Multiplier.
Figure 332. Disconnecting the Electron Multiplier Cables
Electrometer
Board Cable
4. Use a T20 Torxhead screwdriver to remove the screws around the large feedthroughs.
5. Replace the large feedthroughs, o-rings and screws.
6. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
CAUTION - ELECTRICAL HAZARD Avoid dropping screws between chassis and side cover
or board. If you drop a screw, find it before powering the instrument on again.
"Shutting Down the TSQ 9610 System"
Dynode
Electron
Multiplier
Electron
Multiplier Plate