Vent Valve Knob and O-Ring

Thermo Scientific
To replace the vent valve knob and o-ring
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering
these components.
1. Shut down your system by following the steps in
page
139. You do not have to remove the ion source cartridge or column. If you are
using hydrogen as a carrier gas, power off the GC.
.
Figure 323. Powering Off the TSQ 9610 Instrument
2. Remove and replace the vent valve knob and o-ring. Be careful not to pinch the o-ring
during installation.
Vacuum Interlock Shield
Vent Valve Knob
3
Advanced Troubleshooting
Replacing Components of the Manifold
Power Switch
Hydrogen Safety Screws
page
TSQ 9610 Hardware Manual
158.