Convectron Gauge and Foreline Adapter with Hose

Thermo Scientific
A convectron gauge measures and controls the pressure in the foreline that is connected to the
turbomolecular pump and rough pump. If the foreline pressure gets too high, convectron
gauge decreases the power going to the turbomolecular pump. The convectron gauge and
foreline adapter are connected to the foreline hose, which is connected to the turbomolecular
pump.
To replace a convectron gauge and foreline adapter with hose
Note See the ISQ and TSQ GC-MS Spare Parts Guide for ordering information.
1. Shut down your system by following the steps in
on
page
139. You do not have to remove the ion source cartridge.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws securing the left hand front
panel to the chassis. One of the screws is located on the interior front panel of the
instrument.
Figure 469. Locating the Left Hand Front Panel Screws
Left Hand Front
Panel
4. Lift the left hand front panel off the instrument.
5. Move the GC away from the TSQ 9610 instrument.
3
Convectron Gauge and Foreline Adapter with Hose
"Shutting Down the TSQ 9610 System"
Screws
TSQ 9610 Hardware Manual
Advanced Troubleshooting
Screw on Front
of Instrument
429