Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
258
TSQ 9610 Hardware Manual
20. Remove the axle bearing from the metal valve body. Replace the axle bearing, if necessary.
Figure 281. Replacing the Axle Bearing
Axle Bearing
Safety
Interlock
21. Pull the valve axle out of the side of the metal valve body. Replace the valve axle, if
necessary.
22. Remove the axle seal from the metal valve body. Replace the axle seal on the valve axle if
necessary.
Figure 282. Replacing the Axle Seal
Axle Seal
Clip
Thermo Scientific