3
Advanced Troubleshooting
Replacing the Power Supplies
378
TSQ 9610 Hardware Manual
9. Use a T20 Torxhead screwdriver to remove the screw connecting the dynode/multiplier
power supply to bottom of the chassis.
Figure 415. Removing the Controller Interface Board Support Bracket
Dynode/Multiplie
r Power Supply
10. Pull the dynode power supply straight up and away from the distribution board on the
bottom of the chassis.
11. Replace the dynode/multiplier power supply and install it by aligning its connector with
the connector on the distribution board beneath it. It will drop into place when it is in the
right position.
Note A new dynode/multiplier power supply may have a right angle bracket attached to
the top of the supply. Remove this bracket before placing the supply into the TSQ 9610
instrument.
Screw
Thermo Scientific