Thermo Scientific TSQ 9610 Hardware Manual page 278

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing Components of the Source Exchange Tool
248
TSQ 9610 Hardware Manual
To replace the components of the source exchange tool:
1. Use a T20 Torxhead screwdriver to remove the three screws at the end of the black
handle.
Figure 262. Removing Source Exchange Tool Screws
Screws
2. Loosen the set screw at the other end of the black handle.
3. Slide the black handle toward the other end of the tool.
4. Use a T10 Torxhead screwdriver to remove the set screw holding the endcap onto the
interior shaft.
5. Remove the endcap.
Figure 263. Removing the Endcap
Endcap
6. Slide the handle off the rest of the tool. Set it aside.
7. Remove the interior shaft from the barrel.
Figure 264. Removing the Interior Shaft
Set Screw
Barrel
Set Screw
Interior Shaft