Thermo Scientific TSQ 9610 Hardware Manual page 420

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Calibration Gas Components
390
TSQ 9610 Hardware Manual
5. Attach a new calibrant reservoir to the calibration gas module.
Figure 428. Replacing the Calibrant Reservoir and Cover
Calibrant
Reservoir
6. Fill the calibrant reservoir of the new module by following the instructions in
the Calibrant Reservoir"
7. Reattach the calibrant reservoir cover.
8. Close the front door of the instrument.
9. If you are using hydrogen as a carrier gas, attach the hydrogen safety screws and restart
your GC by following the instructions in
on
page
82.
"Restarting the System"
Calibrant
Reservoir
Cover
"Refilling
on
page
158.