Reassembling the ExtractaBrite Ion Source Cartridge

Thermo Scientific
10. Place the locking ring on top of the repeller spring so that the repeller protrudes through
the center hole on the locking ring. The hooks on the sleeve fit between the larger gaps on
the locking ring.
Note Do not twist and lock the locking ring on the sleeve at this time.
Figure 59. Inserting the Locking Ring into the Sleeve
Larger Gaps on
Locking Ring
1. Place the ion volume-repeller insulator on the repeller and hold it in place with the
repeller nut. Set it aside for now.
2. Insert the ion cartridge sleeve into the source holder.
Figure 60. Inserting the Sleeve into the Source Holder
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Hooks on Sleeve
TSQ 9610 Hardware Manual
63