Cleaning The Filters - Thermo Scientific TSQ 9610 Hardware Manual

Mass spectrometer
Table of Contents

Advertisement

1
Performing Routine Maintenance

Cleaning the Filters

Cleaning the Filters
106
TSQ 9610 Hardware Manual
To clean the filters and fan filter on the TSQ 9610 instrument:
Note If your filters wear out or get damaged, refer to the ISQ and TSQ GC-MS Spare
Parts Guide for information about ordering new ones.
1. Open the front door of the instrument.
2. If your instrument has an optional dust filter, use a T20 Torxhead screwdriver to remove
the two M4 screws holding the dust filter frame to the front interior panel.
3. Use a T20 Torxhead screwdriver to remove the six M3 screws holding the EMI filter to
the front interior panel.
Figure 104. Cleaning the Filters
Filters
Turbo Cooling Fan
Filter
Thermo Scientific