16. If you need to replace the lens heater block, remove the screw holding the lens plate and
springs to the lens heater block.
Figure 178. Replacing the Lens Heater Block
17. Remove and replace the lens heater block.
18. Attach the lens plate and springs to the lens heater block.
19. Attach the source heater block.
20. Reattach the lens/source heater to the new source heater block and set them next to the
rest of the ion source.
Figure 179. Reattaching the Lens/Source Heater
3
Advanced Troubleshooting
Replacing Components of the Ion Source
Lens Heater Block
Lens Plate
and Spring
Screw
Lens/Source
Heater
Screws and Cone
Springs