Thermo Scientific
Figure 72. Inserting the Ion Volume and Insulator into the Ion Cartridge Sleeve
5. Insert the filament shield into the filament insulator.
Figure 73. Inserting the Filament Shield into the Filament Insulator
6. Insert the filament shield and insulator assembly into the ion cartridge sleeve. The shield
is keyed to the sleeve for proper insertion.
Figure 74. Inserting the Shield and Insulator Assembly into the Ion Cartridge Sleeve
7. Insert the filament into the ion cartridge sleeve. It will only align with the grooves on the
filament insulator. Press down on the filament so that it bottoms out on the insulator.
Figure 75. Inserting the Filament into the Ion Cartridge Sleeve
8. Put the spring inside the insulating cap.
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
TSQ 9610 Hardware Manual
71