3
Advanced Troubleshooting
Replacing the Calibration Gas Components
384
TSQ 9610 Hardware Manual
10. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
loosen the screw holding the lens driver board to the chassis.
Tip The bracket has a white diagram describing the location of this screw. It is
approximately two inches from the top of the board directly under the white label.
Figure 422. Removing the Lens Driver Board
11. Loosening the screw will separate the board from the instrument. The screw is captured,
so do not attempt to completely remove it from the board and bracket.
12. Lift the lens driver board out of the instrument. Be careful as space is limited.
CAUTION - INSTRUMENT DAMAGE Limit handling of the boards as much as possible.
Do not touch any board components. Hold the board by the sheet metal bracket only.
Lens Driver Board Screw