Figure 488. Removing the Controller Interface Board and PC Communication Board
PC
Communication
Controller
Interface Board
12. Remove the foreline gas tube from the CI reagent gas flow module.
13. Remove the 1/16 in. stainless steel tubing on the output connector of the CI reagent gas
flow module.
14. Use a T20 Torxhead screwdriver to detach the two screws holding the CI reagent gas flow
module to the chassis
Figure 489. Installing the CI Reagent Gas Flow Module
CI Output
Connector
Eth
t C bl
Screws
4
Upgrade Equipment
CI Reagent Gas Flow Module
CI Reagent Gas
Flow Module
Foreline Gas
Tube