Thermo Scientific TSQ 9610 Hardware Manual page 352

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
322
TSQ 9610 Hardware Manual
Figure 357. Disconnecting the Rod Driver Board 7-Pin Connector
Note Be careful not to bend the connector on the rod driver board on the pins on the RF
board.
10. Replace the rod driver board if necessary. Check the pin connections for accuracy when
you connect the board.
Figure 358. Replacing the Rod Driver Board
11. Attach the rod driver board to the new RF board.
12. Reattach the rod driver board's 40-pin and 7-pin connectors to the RF board.
13. Reattach the three screws and two standoffs.
7-Pin Connector
Thermo Scientific