Thermo Scientific TSQ 9610 Hardware Manual page 388

Mass spectrometer
Table of Contents

Advertisement

3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9610 Instrument
358
TSQ 9610 Hardware Manual
24. Use a T20 Torxhead screwdriver to remove the two screws holding the power module to
the instrument.
Figure 397. Removing the Power Module Screws
Screws
25. Pull the power supply out of the top of the instrument.
26. Remove the screw that connects the dynode/multiplier power supply to the bottom of the
chassis.
Thermo Scientific

Advertisement