62
TSQ 9610 Hardware Manual
8. Insert the large flat end of the repeller so that it rests on top of the ion volume.
Figure 57. Inserting the Repeller into the Source Sleeve
9. Slide the repeller spring onto the repeller.
Figure 58. Inserting the Repeller Spring into the Sleeve
Large Flat
End of
Repeller
Repeller
Spring
Thermo Scientific