5. With the longer teeth of lens 1 facing down toward lens 2, align the larger metal section
of lens 1 with the sleeve window and let it fall into place.
Figure 63. Inserting Lens 1 into the Source Sleeve
Sleeve
Window
6. Insert the ion volume with the handles fitting into the notches of the sleeve. Make sure
the ion volume is firmly seated into the gap on lens 1. You may need to rotate lens 1
slightly to make the ion volume fit correctly.
Figure 64. Inserting the Ion Volume into the Source Sleeve
Larger Metal
Section
Handles Fit
Here
1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Longer Teeth
of Lens 1
Ion Volume
Handles
Lens 1 Gap
TSQ 9610 Hardware Manual
65